| Specification | Data |
|---|---|
| Lamp Source | Tungsten Halogen |
| Measurement Range | 0-3.3 OD |
| Reproducibility/Precision | 0.25% at 1.0 OD at 450 nm |
| Accuracy | 0.5% at 1.0 OD at 450 nm |
| Standard Filters | 405, 450, 492, 620 nm |
| Wavelength Selection | Filters |
| Wavelength Range | 400-750nm |
| Plate Types | flat, round and v-bottomed 96-well plates |
| Data Output | Export to PC |
| Detection System | Single channel silicon photodiode |
| Reading Speed | 25 seconds |
| Software | Galapagos Software |
| Shaking | None |
| Temperature Control | None |
| Linearity | 0.25% and 0.0025 OD from 0.1 -2.5 OD at 492 nm |
| Computer Interface | USB 2.0 |
| Optical System | Filters |
| Power Requirement | External power supply: 100-240VAC; current rating 50/60 Hz, 1.5 A |
| Validation | Optional 2010 Checkplate with ArSWEP software |
| Wavelength Reproducibility (+/-) | |
| Wavelength Accuracy (+/-) | |
| Dimensions (WxDxH) | 31.5 x 18.2 x 43.5 cm |
| Weight | 6.6 kg |
| Mode of Operation | PC Control |
| Analysis Software | Galapagos Software |
Galapagos for EZ Read Software:
- Create new plate layouts
- Add multiple assays to a plate
- Add multiple plates per analysis
- Easy to create new methods.
- Select only the data you want for data reporting
- Windows 7 & 8 compatibility as well as older Windows OS from XP.
- So easy you won’t need to export your data into Excel.
EZ Read 400 ELISA and EZ Read 400 Research
- Measures flat, round and v-bottomed 96-well microplates and stripwells with standard ANSI SBS footprints are measured
- Accurate measurements ensured by automatic lamp calibration prior to each reading
- USB 2.0 PC connection
- Interference filters for wavelength selection
- Visible range (400-750nm)
Two versions for specific applications requirements:
Biochrom EZ Read 400 ELISA microplate reader comes with 4 filters that can be used to measure most ELISA assays:
|
ELISA Assay Substrate |
Filter |
| NPP, ABTS | 405nm |
| OPD | 492nm |
| Slow TMB, Turbo TMB, Ultra TMB | 450nm |
| Reference wavelength | 620nm |
Biochrom EZ Read 400 Research microplate reader comes with 8 filters that can be used to measure ELISA, total protein and cell proliferation assays.
| ELISA Assay Substrate | Fillter |
| NPP, ABTS | 405nm |
| OPD | 492nm |
| Slow TMB, Turbo TMB, Ultra TMB | 450nm |
| Reference wavelength | 620nm |
| Total Protein Assays | |
| Bradford | 595nm |
| BCA | 562nm |
| Lowry | 650nm |
| Cell Proliferation/Viability Assays | |
| MTT | 570nm |
| XTT | 492nm |
| Cell Density | 620nm |