Overview:Â
behr distillation units of the S series are the ideal supplement to the behr InKjel digestion systems. Depending on requirements, the user can choose between five steam distillation units. These are identical in the basic construction but differ in terms of operating comfort and degree of automation. The top model behr S 5 is prepared for working with an external titration device.
- Exemplary safety and reliability
- Robust and insensitive casing made of polyurethane
- Distillation time approx. 3 min. per sample
- Detection limit 0.2 mg N
- Recovery rate 99.5%
- Reproducibility ± 1%
- Adjustable steam generating capacity (10% – 100%)
- Very easy, menu-guided operation with a single operation element (the behr one-button operation)
- Programmable reaction time
- Programmable distillation time
- USB interface
- Fill-level monitoring for canister set
- Practical quick clamping device which the user can operate even “with his eyes closed”
- Performance capacity and speed
- Practical and user friendly
- Exemplary safety and reliability
- Robust and insensitive casing made of polyurethane
- Practical quick clamping device which the user can operate even “with his eyes closed”
- Extremely simple, menu-controlled operation of the programmable units via
- one single operating element (behr one-button operation).
- Menu navigation in country language
- Models S 2 to S 5 have a separate rinsing program
Technical data:Â
Item number | S1 | S2 | S3 | S4 |
---|---|---|---|---|
Packaging quantity | 1 | |||
Width (in cm) | 42.00 | |||
Depth (in cm) | 34.50 | |||
Height (in cm) | 70.00 | |||
Weight (in kg) | 32.00 | 34.00 | 35.00 | 35.00 |
Controller | Nein | |||
Power consumption (in W) | 1700 | |||
Current consumption (in A) | 9 | |||
Programs | 1 | 1 | 10 | 99 |
Cooling water consumption (in l/min) | 5 | |||
max. distillation time (in min/sample) | 2-4 | approx. 2-4 min. per sample | approx. 2-4 min. per sample | approx. 2-4 min. per sample |
Supply container | any size; recommended: behrotest® canister set KAS | any size; recommended: behrotest® can set KAS | any size; recommended: behrotest® can set KAS | any size; recommended: behrotest® can set KAS |
Benefits:
Safe and high-performance steam distillation units
- Door contact safety switch that is automatically triggered in the event of overload or short circuit
- Mechanical overpressure safety valve against excessive pressure in the steam generator
- Vessel monitoring (no distillation possible if no vessel has been inserted)
- Door contact switch that automatically deactivates the distillation unit if the door is open
- Resettable overtemperature thermostat (in the case of water shortage in the steam generator)
- Cooling circuit monitoring by means of pressure switch
- Temperature-regulated steam heating phase and pressure control via solenoid valve
behr steam distillation units are not just high performance units and reliable partners in your daily laboratory routine. An essential factor in the development and design of the devices was also the safety of the user