Measures octane number via near-infrared (NIR) transmission spectroscopy utilizing 14 near-infrared emitting diodes with narrow bandpass filters, a silicon detector system, and a fully integrated microprocessor. Simple octane number determination requires three easy steps: sampling a background signal, acquiring two absorption spectra of the gas sample, and then acquiring a second background signal.
Features:
- Test results equivalent to ASTM D2699 and D2700 test methods
- Measures all grades of unleaded gasoline and ethanol blended gasoline
- Test results equivalent to ASTM D613 for Cetane Number of diesel fuels (Optional with K88612)
- Displays results in 20 seconds
- Directly measures octane number for {(R+M)}/2, RON and MON
- Optional feature for cetane number determination of diesel fuels
- Includes RS-232 output, built-in printer and LCD display
- Results traceable to official knock engine laboratory
- GPS model available for use with GPS locator accessory Safety Features Out of Temperature Range Warning: Analyzer displays Out of Range Warning Message when instrument in being used outside of its standard operating temperature range. Either above 38°C or below 7°C.
- Out of Calibration Range Warning: Analyzer displays “Too High” or “Too
- Low” message when measurement reading is out of the instruments calibration range.
- Bad Curve Warning: Analyzer warns user when light protective lid is not on during testing. External light source will greatly disrupt results.
Specifications:
Accuracy and repeatability equivalent to ASTM approved CFR engines test methods (ASTM D2699, D2700)
Sample Holder: Sealed, cubical glass container (75mm optical path length)
Sample Volume: 8 Ounces (approx. 225 mL)
Operating Temperature Range: 7°C – 38°C – Pre-calibrated for unleaded & ethanol-blended gasoline (Analyzer can be calibrated for up to 8 additional fuel types.) Battery operated (6 AA batteries)