High Performance Analysis
* High sensitivity and response speed
* High UV sensitivity and accurate detection of short-wave elements in deep UV such as C, P, S, B, As, N
* Vacuum chamber integration
Stable And Reliable Operating Capability
* Interference subtraction technology
* Drift correction technology
* Easily to use
Intelligently Working Program
* Prefabricated work curve with different materials and grades
* Optimal test condition be automatically selected
* Scope of analysis be drawn or extended by user
Advanced Analytical Software
* Fast analysis speed, takes 20s for one test
* According to different analysis materials, setting proper preburning and measuring time to achieve best analyzing effect
Optical System |
Paschen-Runge optical system |
High resolution PMT detectors | |
Wavelength:175-520 nm | |
Curvature radius:750mm | |
Constant temperature | |
Spark Stand | Open spark stand reduces the gas consumption |
Sample clamp :Customized | |
Four-way argon flushing design | |
Easy to clean and maintain | |
Spark Source |
Programmable digitalized source |
Maximum discharge frequency: 100-1000Hz, | |
Maximum discharge current: 400A | |
Spark Excitation Pulse: 20-230V | |
Aperture: 12mm | |
Argon |
Purity: 99.999% |
Import pressure: 0.5 MPa | |
Excitation flow:3.5L/min Standby flow: 0.1L/min. |
|
Power Supply | 50Hz、220V±1% |
Working Environment: | 10℃ to 30℃ |
Size | L1130mm*W750mm*H920mm |
Weight | Approx 280kg |
JB-750 optical emission spectrometer breakthrough new versions, this analyzer sets a new standard for R&D and process/quality control. enables the analysis of Fe, Al, Cu, Zn, Ni, Pb, Mg etc. bases materials.this version provides unbeatable precision, flexibility, and stability for specialized R&D laboratories.